EXPERIMENT #5
Introduction
to the Ion Chromatograph: Calibration
and Blanks
OVERVIEW
The
purpose of this lab is to become familiar with the Dionex
DX-500 Ion Chromatograph. This
instrument is used to perform the analyses for several other lab exercises
(filter samples, impactors, denuders, precipitation
sampling). The first step in running the
ion chromatograph (IC) is to prepare standards of known concentration and
construct a calibration curve. In this
exercise each group will prepare standards and their own calibration
curve.
Equipment Needed
Lab
Notebook, Safety Glasses, Lab Coat
Latex
gloves
Ion
Chromatograph
Balance
Weighing
boats
Teflon
spatula
Deionized water from water system
Wash
bottle with deionized water
1
l volumetric flask
100
ml volumetric flask
Pipettes
and pipette tips
Inorganic
Salts: Sodium Chloride, Sodium Nitrate,
Sodium Sulfate
Autosampler vials
Plastic
bottles for storing standards
Single
component standards (provided by teaching assistant)
Mystery
Sample (provided by teaching assistant)
Procedure
Preparing
a blank sample
You
should probably wear latex gloves for any of the procedures where you might
contact the samples or standards. While
the gloves are not particularly clean, your hands probably have lots of salts
on them, especially right after lunch.
Wash the gloves with DI water after putting them on.
In
order to check that the IC is operating properly, we will run a blank sample
and single component standards. The
teaching assistant will have started the IC in the morning so it should be
ready to work by the lab period.
The
autosampler vials are small plastic vials with
screw-type caps. The caps have a septum
in them. As a group we will prepare one
blank sample vial, and one single component standard vial each for chloride,
nitrate, and sulfate. Fill each vial
with approximately 0.6 ml of the appropriate solution (Nanopure
water or a single component standard) and place a cap on it. To avoid confusion label each vial
appropriately with a permanent marker.
Preparation
of Standards
To
save time during this laboratory session, we will only use four standards to
prepare a calibration curve. For
research analytical work we generally use six or more standards. The standards will contain known amounts of
nitrate, sulfate and chloride.
A
stock solution is a convenient tool for preparing the standards. A reasonable amount of salt is weighed out
for a 1 liter stock solution which is then diluted to the concentrations you
need. The concentration of the stock
solution depends on what levels you expect to find in your samples. Since you probably don't have a good idea of
what these concentrations are right now (and won't have time in this laboratory
period to experiment too much), choose a stock solution of :
0.01 Normal
Na2SO4
0.01 Normal NaNO3
0.005 Normal NaCl
The lowest of the four standards you should prepare for this
exercise is 5 µN Cl- / 10 µN NO3-
/ 10 µN SO42-, and the highest should be 10 times
those amounts. Choose two intermediate
values for your other two standards.
Prepare
the stock solution by weighing the amount of salts you need and then adding DI
water to fill a 1 liter volumetric flask.
You can make the dilutions using a 100 ml volumetric flask and the
micropipettes.
Place
approximately 0.6 ml of each standard into an autosampler
vial. Do the same for the "mystery
sample" and a blank sample of Nanopure water. Label each vial appropriately and give them
to the TA.
Running
the IC
The
teaching assistant will show you how to place the autosampler
vials into the autosampler. In order to save time, for this lab exercise
the teaching assistant will set up the method (instructions for controlling the
IC) and the schedule (list of samples the autosampler
will inject into the column). Setting
these up isn't very difficult, but the software takes some getting used to.
Constructing
a calibration curve
Although
the Dionex data processing software can prepare
calibration curves, for this lab each group will construct their own
calibration curves. You need to prepare
a calibration curve for each ion in the standards you've prepared.
For
each chromatogram, the IC software determines the area under each peak. The teaching assistant will print out a
report for each chromatogram. The report
shows a picture of the chromatogram and lots of statistics on the peaks, including
area.
The
calibration curve is then constructed by constructing a linear regression
between the areas of the peaks and the concentrations of the standards. The
area is the x-coordinate and the concentration is the y-coordinate.
Once
you have the calibration plot and the linear regression between area and
concentration, you can then use the equation to find the ion concentrations of
any sample from its peak areas. This is
the technique needed to find the "mystery sample" concentration.
Discussion/Lab Write-Up
Plot
the calibration data for your standards and find the equation for the linear
regression between concentration and area for each anion. Use the relationship you've found above to
calculate the concentrations of your standards (i.e., plug the peak areas for
your standards back into the calibration equations to predict
concentrations). Are they close to their
nominal values?
What
were the ion concentrations in the "mystery sample"?
Comment
on your calibration curve. Discuss the
intercept value. Would more points
improve your curve? Discuss the scatter
of your standard concentrations about the calibration curve.
What
ion concentrations did you obtain from the peak areas found for your blank
sample?
Discuss
how errors in preparing standards would affect the concentrations of samples
you would measure. What if you were making your standards with the 100 ml
volumetric flask and you accidentally overfilled the flask on the first
standard? Would you be able to get a
decent calibration curve if you just overfilled the flask the same amount for
the other standards? Would the
concentrations you get for samples be accurate?
Suppose an unidentified peak showed up in your chromatograms. How might you decide what this peak was?