EXPERIMENT #1
Pipette Calibration, pH Measurement and
Statistics
OVERVIEW
The
purpose of this experiment is to become familiar with the lab by making some
simple measurements with some of the tools we will use for other labs. The statistical concepts we have discussed in
lecture (mean value, standard deviation, confidence intervals, etc.) will be
applied to the calibration of micropipettes and to the pH measurement of several
samples.
Equipment Needed
Lab
Notebook, Safety Glasses, Lab Coat
Balance
Weighing boats
Teflon spatula
Deionized
water from water purification system
Thermometer
pH Probe
and VWR Symphony pH Meter
Wash bottle with deionized
water
100 ml volumetric flask
Glass Funnel
Micropipettes and pipette tips
Test tubes and rack for holding
samples
Beaker of salt (provided by teaching
assistant)
Dilute sulfuric acid sample (provided
by teaching assistant)
Cloud/fog samples (provided by
teaching assistant)
Table of density of water (provided
by teaching assistant)
Procedure
Calibration
of micropipettes
The Finnpipette
micropipettes used in this lab are a convenient way to measure and dispense
small volumes of liquids. Disposable
polyethylene tips are used with these pipettes.
The teaching assistant will demonstrate how to attach the tips, how to
pick up liquids and dispense them, and how to discard the used tips. In later lab exercises it will be important
to be able to use these pipettes correctly in order to properly measure
solutions and prevent contamination of samples.
Check that the analytical balance
you will use is level. Set the 40 - 200
µl pipette to 200 µl and use it to pick up a clean
tip. Place a clean weighing boat on the
balance and tare the reading to zero (depending on the balance you are using,
you may have to open the glass doors in order to place things on the balance;
it's a good idea to close the doors when taking readings to prevent
interference from air currents). Add 200
µl of deionized water (DI water) and record the reading. Repeat this for a total of five measurements
(either remove the weighing boat, empty it, dry it, and re-tare it, or continue
adding water to the weighing boat and re-tare it in between measurements --
which method do you think is more accurate?).
Set the 200 - 1000 µl pipette to 1000 µl and record 5 measurements. Record the temperature of the DI water. The teaching assistant will have a chart of
the density of water as a function of temperature. Use this chart and record the density of the
water you have used for these measurements.
Calibrate pH probe
The teaching assistant will demonstrate
how to connect the pH probes to the portable Orion pH meters. In order to calibrate the pH probe you will
use buffer solutions. We will use pH
4.01 and pH 7 buffers. Transfer
approximately 1 ml of the pH 4.01 buffer into a test tube and place it in the
test tube rack. Do the same for a second
tube with pH 7 buffer.
First rinse the tip of the probe
with DI water using your wash bottle.
Gently touch the end of the probe against the plastic beaker to remove
any water adhering to the tip. Rinse the
probe with ~500 µl of the pH 7 buffer.
Place the probe into the test tube with the 1 ml of pH 7 buffer
solution.
In order to start the calibration,
first set the temperature. Turn on the
pH meter. Press the setup key then press
the DOWN ARROW 4 times to reach the temperature setting. Press the RIGHT ARROW key
(has three boxes with arrows) twice to access the temperature setting. Use the DIGIT CHANGE KEY (left of the arrows)
to select each digit and the UP/DOWN ARROWS to adjust the temperature digits as
needed to match the temperature of the water you measured above. Press the RIGHT ARROW then the DOWN ARROW to
exit the temperature set mode, then press the MEASURE key to exit the
setup. The temperature you set will be
displayed in the upper left corner of the display.
Place the pH electrode into the pH 7
buffer, then press the CALIBRATE button.
The meter should recognize the buffer and set the display to the
appropriate value. Press the CALIBRATE
key to accept the autobuffer recognition as the first
calibration point and start the second calibration point (4.01) when prompted
for the second buffer. After the second
buffer is autorecognized, press the MEASURE key to
exit calibration mode and return to measurement mode. Before returning to measurement mode, the
electrode calibration slope will be displayed.
Record this in your lab book. You
should find a slope close to 100%. The slope should be at least > 90%. If not, calibration should be repeated until
a high enough slope is obtained. If this
takes more than two tries, consult the TA.
If something went wrong with the
calibration, an error code may appear and you will have to perform the calibration
again. If the problem persists, consult
the TA.
Checking the calibration
periodically is a good idea. If one of
the buffer solutions does not read within a couple one-hundredths of the
calibration value, you should recalibrate.
In between measurements (or when the pH portion of this experiment is
completed), leave the probe soaking in the pH 4.01 buffer. Do not leave the probe for an extended period
of time in DI water.
pH
Measurement of the Diluted Acid Solution
Now that the pH probe has been
calibrated, you can measure the pH of the diluted acid solution the teaching assistant
has provided. Rinse the pH probe with
0.5 ml of the acid solution and then place it in a test tube with approximately
1 ml of the acid solution. Take three
measurements of the pH of this solution.
Use the same tube with a fresh portion of acid each time, rinsing the
probe with the dilute acid solution between each measurement.
You now want to make three
measurements of a 1/10 dilution of this acid solution. Prepare enough diluted acid (mix well!) to
make three pH measurements, remembering you need to rinse the electrode with
solution prior to each measurement. Make
the pH measurements and record the results in your lab notebook (you should be
recording all procedures and results there).
pH
of Deionized Water
Measure the pH of the DI water
(three times) and record the result in your lab notebook.
The teaching assistant has provided
a bottle of an inorganic salt. The
molecular weight of the salt is marked on the bottle. Each group should make a 0.1 M solution of
this salt. When you're ready to make the
solution, use one of the spatulas to transfer the salt from the bottle. Remember not to put any chemical back into
the original salt container, and clean up anything you may spill on the
balance. The teaching assistant can show
you a way to steady your hand as you transfer chemicals from the spatula into the
weighing boat.
Recheck your calibration using fresh
pH 7 and pH 4.01 buffers. Measure the pH
of the solution you've made and record the results in your lab notebook.
pH
of Cloud/fog Samples
The teaching assistant will provide
you with actual cloud/fog samples collected from two different locations. Measure the pH of each sample three times,
using the same procedures you've used above with 0.5 ml for rinses and a fresh
1ml for the measurements.
Data Exchange
Before you leave the laboratory, be
sure to review the list of topics below to discuss in your lab write-up. Exchange data with other groups as needed.
Discussion/Lab Write-Up
Begin
this, and each lab writeup, with a brief summary of
the experiment in your own words. State
the purpose of the experiment. Answer
each of the questions below and address the discussion points. Consider these questions/discussion items as
a starting point. Expand your
analysis/discussion to include other relevant or interesting points as you see
fit.
Statistical Evaluation
Using the statistical methods we've
discussed in class and the material in Chapters 2-4 of the Fundamentals of Analytical Chemistry text by Skoog,
West and Holler, answer the following questions:
1.
What are the means and
the 95% confidence limits for the pipette measurements?
2.
What is the 95%
confidence limit of your pH measurements for cloud/fog samples with pH <
4? Use pooled statistics for data from
your group and other AT 560 groups to obtain a pooled standard deviation.
3.
Are the pH measurements
of the two cloud/fog samples you measured different at the 95% confidence level? Be sure to perform a hypothesis test; don’t
just check whether the confidence intervals overlap.
4.
Compare your measurements
of the cloudwater sample pH values with those
obtained by other lab groups. Are your
results statistically different?
Discussion
o
Comment on your experience with any
difficulties measuring the pH of individual samples.
o
What effect did the salt addition have on
your ability to measure the deionized water pH value?
o
Compare the pH measurements of the
original and diluted acid solutions. Is
the pH difference what you expected?